Problem/Question
Label does not stop at the correct position.
Labels are misaligned.
Skipping Labels.
Applicable To
S84NX / S86NX
Resolution/Answer
Perform the auto-calibration for the selected media sensor.
[Auto-calibration] is not available if you have selected [Linerless] in the [Print Mode].
Gap + I-Mark
Adjusts both the Gap sensor and I-mark sensor.
Gap
Adjusts the Gap sensor.
I-Mark
Adjusts the I-mark sensor.
The procedure to auto-calibrate the media sensor is as follows:
1.
|
Open the top cover.
|
2.
|
Turn the head lock lever clockwise to unlock the print head, and then pull the feed lock latch to unlock the feed roller and media sensor assembly.
|
3.
|
Load the media.
|
Remove the label from the liner and set the liner so the I-mark is not above the I-mark sensor.

4.
|
Turn the head lock lever counterclockwise to lock the print head, and then press the feed roller and media sensor assembly down until the feed lock latch is locked.
|
|
|
5.
|
Close the top cover.
|
6.
|
Press the
![]() ![]() |
7.
|
Press the
![]() ![]() |
8.
|
When the confirmation screen appears, press the
![]() |
9.
|
The result of the sensor adjustment is shown. To exit the adjustment, press the
![]() |
10.
|
Set to Offline mode. Press the
![]() |
03-80162-EN
Comments
0 comments
Please sign in to leave a comment.